Preclinical PET Systems: Comparative Analysis and Technical Specification Guide

Sistemas PET Pré-clínicos: Análise Comparativa e Guia de Especificação Técnica

Autores/as

  • Lívia Regina Medeiros Pereira UFMG
  • Rodrigo Modesto Gadelha Gontijo UFMG , CMMG , EBSERH , Centro de Desenvolvimento da Tecnologia Nuclear image/svg+xml

DOI:

https://doi.org/10.70745/bjrtr.v1.232

Palabras clave:

positron emission tomography (PET), Preclinical Research, Technical Specifications

Resumen

Positron Emission Tomography (PET) is a high-sensitivity and high-specificity molecular imaging technique, fundamental for preclinical research in small animals. Since the 1990s, technological advances have generated a variety of equipment with configurations that directly impact data quality, creating a heterogeneous landscape for researchers. The aim of this work is to provide a standardized comparative framework by performing a systematic analysis of technical specifications to assist laboratory heads and users in selecting the most appropriate equipment according to their specific research needs. The methodology involved a systematic review of current literature and manufacturer technical portfolios. Data were organized into a comparative framework to analyze parameters such as field of view; spatial resolution; sensitivity; detector crystal type and size; detection technology; reconstruction algorithms; Depth-of-Interaction capability; and multimodal integration. The analysis revealed that the optimization of PET systems is a complex balance between spatial resolution and sensitivity. It is important to emphasize that the in-depth analysis of certain systems is constrained by manufacturers' intellectual property, reinforcing the necessity of technical guides to compare publicly declared parameters. This study offers a solid comparative basis as a foundation for institutional decision-making, ensuring the acquisition of high-quality and precise data for molecular imaging analysis. By establishing a standardized selection framework, this work facilitates the optimization of resources and the advancement of preclinical research strategies.

Biografía del autor/a

  • Lívia Regina Medeiros Pereira, UFMG

    Tecnóloga em Radiologia

  • Rodrigo Modesto Gadelha Gontijo, UFMG, CMMG, EBSERH, Centro de Desenvolvimento da Tecnologia Nuclear

    Doutor em Ciência e Tecnologia das Radiações, Minerais e Materiais [CDTN/CNEN]. Mestre em Radioproteção e Dosimetria com ênfase em Biofísica das Radiações [IRD/CNEN]. Tecnólogo em Radiologia [CEFET-MG]. Professor Associado UFMG e Adjunto CMMG; Tecnólogo em Radiologia HC-UFMG / EBSERH. 

Referencias

1. Cherry SR. Fundamentals of positron emission tomography and applications in preclinical drug development. J Clin Pharmacol. 2001;41(5):482-491. DOI: https://doi.org/10.1177/00912700122010357

2. Khalil M. Small Animal micro-PET imaging: an overview. Egypt J Nucl Med. 2017;14(14):8-27. DOI: https://doi.org/10.21608/egyjnm.2017.5435

3. Amirrashedi M, Zaidi H, Ay MR. Advances in preclinical PET instrumentation. PET Clin. 2020 Oct;15(4):403-426. DOI: https://doi.org/10.1016/j.cpet.2020.06.003

4. Gonzalez-Montoro A, Ullah MN, Levin CS. Advances in detector instrumentation for PET. J Nucl Med. 2022;63(8):1138-1144.Pichler BJ, Wehrl HF, Judenhofer MS. Latest advances in molecular imaging instrumentation. J Nucl Med. 2008;49 Suppl 2:5S-23S. DOI: https://doi.org/10.2967/jnumed.121.262509

5. Pajak MZ, et al. NEMA NU4-2008 performance evaluation of Albira: A two-ring small-animal PET system using continuous LYSO crystals. Open Med J. 2016;3(1):12-26. DOI: https://doi.org/10.2174/1874220301603010012

6. Sánchez F, et al. ALBIRA: a small animal PET∕SPECT∕CT imaging system. Med Phys. 2013;40(5):051906. DOI: https://doi.org/10.1118/1.4800798

7. Doss KKM, et al. Performance Evaluation of a PET of 7T Bruker Micro-PET/MR Based on NEMA NU 4-2008 Standards. Electronics. 2022;11(19):2975. DOI: https://doi.org/10.3390/electronics11192975

8. Emvalomenos G, et al. Performance evaluation of a PET insert for preclinical MRI in stand-alone PET and simultaneous PET-MRI modes. EJNMMI Phys. 2021;8(1):68. DOI: https://doi.org/10.1186/s40658-021-00415-1

9. Ko GB, et al. Evaluation of a silicon photomultiplier PET insert for simultaneous PET and MR imaging. Med Phys. 2016;43(1):72. DOI: https://doi.org/10.1118/1.4937784

10. Belcari N, et al. NEMA NU-4 performance evaluation of the IRIS PET/CT preclinical scanner. IEEE Trans Radiat Plasma Med Sci. 2017;1(4):301-309. DOI: https://doi.org/10.1109/TRPMS.2017.2707300

11. Boisson F, et al. Imaging capabilities of the IRIS and IRIS XL-260 PET/CT systems for high-throughput imaging. IEEE Trans Radiat Plasma Med Sci. 2023;7(4):326-332. DOI: https://doi.org/10.1109/TRPMS.2022.3222565

12. Panetta D, et al. Performance evaluation of the CT component of the IRIS PET/CT preclinical tomograph. Nucl Instrum Methods Phys Res A. 2016;805:135-144. DOI: https://doi.org/10.1016/j.nima.2015.08.044

13. Szanda I, et al. National Electrical Manufacturers Association NU-4 performance evaluation of the PET component of the NanoPET/CT preclinical PET/CT scanner. J Nucl Med. 2011;52(11):1741-1747. DOI: https://doi.org/10.2967/jnumed.111.088260

14. Nagy K, et al. Performance evaluation of the small-animal nanoScan PET/MRI system. J Nucl Med. 2013;54(10):1825-1832. DOI: https://doi.org/10.2967/jnumed.112.119065

15. Goorden MC, et al. VECTor: a preclinical imaging system for simultaneous submillimeter SPECT and PET. J Nucl Med. 2013;54(2):306-312. DOI: https://doi.org/10.2967/jnumed.112.109538

16. Beekman FJ, et al. Positron range-free and multi-isotope tomography of positron emitters. Phys Med Biol. 2021;66(6):065011. DOI: https://doi.org/10.1088/1361-6560/abe5fc

17. Krishnamoorthy S, et al. Performance evaluation of the MOLECUBES β-CUBE-a high spatial resolution and high sensitivity small animal PET scanner utilizing monolithic LYSO scintillation detectors. Phys Med Biol. 2018;63(15):155013. DOI: https://doi.org/10.1088/1361-6560/aacec3

18. Kuntner C, et al. Assessing PET performance, image quality, and attenuation correction in the presence of RF coils within a 9.4 T MRI for preclinical simultaneous imaging. EJNMMI Phys. 2025;12(1):63. DOI: https://doi.org/10.1186/s40658-025-00771-2

19. Cabello J, Ziegler SI. Advances in PET/MR instrumentation and image reconstruction. Br J Radiol. 2018;91(1081):20160363. DOI: https://doi.org/10.1259/bjr.20160363

20. Raylman RR, et al. Simultaneous MRI and PET imaging of a rat brain. Phys Med Biol. 2006;51(24):6371-6379. DOI: https://doi.org/10.1088/0031-9155/51/24/006

21. Gu Z, et al. Performance evaluation of G8, a high-sensitivity benchtop preclinical PET/CT tomograph. J Nucl Med. 2019;60(1):142-149. DOI: https://doi.org/10.2967/jnumed.118.208827

22. Wang L, et al. Performance evaluation of the Trans-PET® BioCaliburn® LH system: a large FOV small-animal PET system. Phys Med Biol. 2015;60(1):137-150. DOI: https://doi.org/10.1088/0031-9155/60/1/137

23. Ribeiro FM, et al. First performance evaluation of easyPET.3D, a high-resolution and cost-effective benchtop preclinical PET scanner. EJNMMI Phys. 2025;12(1):48. DOI: https://doi.org/10.1186/s40658-025-00757-0

24. Efthimiou N, et al. Influence of multiple animal scanning on Image Quality for the sedecal SuperArgus2R preclinical PET scanner. Front Phys. 2021;8. DOI: https://doi.org/10.3389/fphy.2020.531662

25. Molinos C, et al. Low-dose imaging in a new preclinical total-body PET/CT scanner. Front Med. 2019;6:88. DOI: https://doi.org/10.3389/fmed.2019.00088

26. Arias-Valcayo F, et al. Reconstruction of multi-animal PET acquisitions with anisotropically variant PSF. Biomed Phys Eng Express. 2023;9(6):065018. DOI: https://doi.org/10.1088/2057-1976/acf936

27. Cherry SR, et al. MicroPET: a high resolution PET scanner for imaging small animals. In: IEEE; 2002.

28. Cutler PD, et al. Design features and performance of a PET system for animal research. J Nucl Med. 1992;33(4):595-604. DOI: https://doi.org/10.1111/j.1748-5827.1992.tb01068.x

29. Lee TS, et al. Initial evaluation of a state-of-the-art commercial preclinical PET/CT scanner. In: IEEE; 2016. DOI: https://doi.org/10.1109/NSSMIC.2016.8069505

30. Maus J, et al. Evaluation of effective spatial resolution in reconstructed PET images. In: Georg Thieme Verlag KG; 2021. DOI: https://doi.org/10.1055/s-0041-1726820

31. Prasad R, Ratib O, Zaidi H. NEMA NU-04-based performance characteristics of the LabPET-8TM small animal PET scanner. Phys Med Biol. 2011;56(20):6649-6664. DOI: https://doi.org/10.1088/0031-9155/56/20/009

Publicado

2026-10-06

Cómo citar

Medeiros Pereira, L. R., & Gadelha Gontijo, R. M. (2026). Preclinical PET Systems: Comparative Analysis and Technical Specification Guide: Sistemas PET Pré-clínicos: Análise Comparativa e Guia de Especificação Técnica. Brazilian Journal of Radiation Technology Research (ISSN 2966-4292), 1(01), 1-12. https://doi.org/10.70745/bjrtr.v1.232

Artículos similares

1-10 de 23

También puede Iniciar una búsqueda de similitud avanzada para este artículo.

Artículos más leídos del mismo autor/a